SDD Modules

KETEK’s Silicon Drift Detectors (SDD) are the state-of-the-art X-ray sensors for the energy range between 0.05 keV and 30 keV.
They are globally used in manifold applications in electron microscopes, XRF benchtop and XRF handheld spectrometers as well as in material sorting systems for mining and recycling. Due to their wide operating temperature range, their excellent energy resolution, and high reliability they are particularly suited for industrial applications.

VITUS H80

100 mm² collimated to 80 mm² X-ray Silicon Drift Detector for XRF – EDX – TXRF Applications

Unique Features

  • Large collimated area
  • Very large solid angle
  • New KETEK CSA improves the energy resolution at all peaking times
  • Power consumption less than 1 W for -40 °C chip temperature at 20 °C heat sink temperature
  • Superior radiation hardness up to a total dose of 5×1012 incoming low energy photons
Available in

Be Window

25 µm Beryllium window with DuraCoat Coating
vacuum encapsulated and ultra stable for many years of operation

Specifications

First amplification stage KETEK CSA
Energy resolution (@ Mn-Kα, -45 °C chip temperature and 1 µs peaking time) ≤ 136 eV
Peak to background > 15,000
Peak to tail > 2,000
Optimal peaking time at max. cooling 1 µs
Absorption depth 450 µm Si
Peak shift stability up to 100 kcps < 1 eV
Max. input countrate 2,000 kcps
Window 25 µm Beryllium
Cooling performance
at +20 °C heat sink temperature
∆T > 95 K
On-chip collimator multilayer

Spectrum

The spectrum has been acquired in KETEK’s standard end qualification test stand with an Fe-55 source. The input count rate has been 50 kcps at a peaking time of 1 μs. The spectrum shows an excellent energy resolution for Mn-Kα and peak-to-background ratio.

Energy Resolution

Energy resolution (Mn-Kα) values for peaking times from 0.1 μs through 8 μs showing good FWHM values even for higher operating temperatures. Depending on the application the best performance can be achieved by an appropriate selection of peaking time and set operating temperature.

Cooling Performance

NEW SDD COOLING TECHNOLOGY

  • Increased efficiency of the thermoelectric cooling (e.g. < 1.2 W at -45 °C sensor temperature)
  • Ultra stable vacuum integrity for typically more than 10 years of operation

Operation Requirements

SDD Voltages and Currents
Ring1 (R1) -20 V ± 5 V 10 μA typ.
RingX (RX) -130 V ± 20 V 10 μA typ.
Back -65 V ± 5 V < 1 nA
Peltier Element 9V 1000 mA max.
General Parameters
Temperature Monitor NTC thermistor 10 kΩ @ 25 °C
Output signal ramped reset type
CSA Parameter
VI/O 3.3 V ± 0.1 V < 1 mA
Vs 2 V ± 0.1 V < 1 mA
Vsss -5 V ± 0.1 V < 1 mA
Output gain 1.6 mV/keV ± 20 %

Pin Assignment

Geometry

Overview VITUS SDD Series

KETEK offers a broad range of high performance Silicon Drift Detectors (SDD) for energy dispersive X-ray spectroscopy. From mature off-the-shelf systems to innovative customized solutions, KETEK provides the right X-ray sensor for your needs across the spectrum of scientific and industrial applications. The VITUS SDD series combined with KETEK’s complete electronics including the pulse processing is a fast and efficient acquisition solution of highest quality.

SDD Systems

Highly optimized components for an unprecedented performance.

SDD Accessories